Thermo Fisher Scientific Ion Selective Electrodes Nitrate User Manual
Page 17

Instruction Manual
Nitrate Ion Electrode
15
Low Slope or
standards contaminated
prepare fresh standards
No Slope
or incorrectly made
ISA not used
use recommended ISA
standard used as ISA
use ISA
electrode exposed
soak electrode in
to interferences
nitrate standard
defective electrode
check electrode operation
air bubble on
remove bubble by
membrane
re-dipping probe
"Incorrect Answer" incorrect scaling
plot millivolts on the linear axis. On the
(but calibration
of semi-log paper
log axis be sure concentration numbers
curve
is
good)
within
each
decade
are
increasing
with
increasing
concentration.
.
incorrect sign
be sure to note sign of millivolt number
correctly
incorrect standards
prepare fresh standards
wrong units used
apply correct conversion factor:
10
-3
M
=
62.0
ppm
as
NO
3
= 14.0 ppm as N
sample carryover
rinse electrodes thoroughly between
samples