Thermo Fisher Scientific Alpha TDS 200 User Manual
Page 29

25
9 RELAYS
The TDS 200 features two SPDT non-powered relays; rated for 6A at 110 VAC, 250 VAC
maximum. When your process exceeds the set parameters of a relay set point, the REL
1 or REL 2 indicator lights up.
To set parameters for relay one and relay two, see Setup programs P1.0 and P2.0.
10 TRANSMITTER FUNCTION
If remote data logging is required, a 4-20 mA current loop can be connected. The current
will be proportional to the TDS value displayed on the panel and according to the
measurement range selected.
TDS Range
Range No., R
Cell Constant
4 mA current
20 mA current
0.00 – 10.00 ppm
1
0.1
0.00 ppm
10.00 ppm
0.0 – 100.0 ppm
2
0.1
0.0 ppm
100.0 ppm
0.0 – 100.0 ppm
3
1
0.0 ppm
100.0 ppm
0 – 1000 ppm
4
1
0 ppm
1000 ppm
0.00 – 5.00 ppt
5
1
0.00 ppt
5.0 ppt
0.00 – 10.00 ppt
6
1
0.00 ppt
10.0 ppt
0.0 – 100.0 ppt
7
1
0.0 ppt
100.0 ppt
0.0 – 100.0 ppt
8
10
0.0 ppt
100.0 ppt
The 4-20 mA current loop can drive a load resistance of no more than 200 Ω.