8 SETUP MODE
8.1 General Information
IMPORTANT: When Setup mode is entered, controller automatically goes into a
“HOLD” mode where the 4-20 mA output freezes and relays are de-activated (if it
was in an activated condition). Upon return to measurement mode, both 4-20mA
output and relay activities resume, depending on settings.
To enter setup mode:
1.
Key in password “022” using method described in section 5.
2.
Press ▲or ▼keys to display various sub-menus shown here.
3.
When a sub-menu item is displayed, press ENTER key to enter that sub-
menu.
4.
Press ▲and ▼keys together (ESCAPE) to leave Setup mode and return to
Measurement mode (Relays and 4-20 mA output resume previous settings).
P1.0