Thermo Fisher Scientific Alpha TDS 190 (In the process of updating the brand guideline & product image) User Manual
Page 29

26
9 RELAYS
The
αlpha TDS 190 features two SPDT non-powered relays; rated for 6A at 110
VAC, 250 VAC maximum. When your process exceeds the set parameters of a
relay set point, the REL 1 or REL 2 indicator will light up.
To set parameters for relay 1 and relay 2, see Setup programs P1.0 and P2.0.
10 TRANSMITTER FUNCTION
If remote data logging is required, a 4-20 mA current loop can be connected. The
current will be proportional to the TDS value displayed on the panel and according
to the measurement range selected.
TDS Range
Range No., R
Cell Constant
4 mA current
20 mA current
0.00 – 10.00 ppm
1
0.1
0.00 ppm
10.00 ppm
0.0 – 100.0 ppm
2
0.1
0.0 ppm
100.0 ppm
0.0 – 100.0 ppm
3
1
0.0 ppm
200.0 ppm
0 – 1000 ppm
4
1
0 ppm
1000 ppm
0.00 – 5.00 ppt
5
1
0.00 ppt
5.0 ppt
0.00 – 10.00 ppt
6
1
0.00 ppt
10.0 ppt
0.0 – 100.0 ppt
7
1
0.0 ppt
100.0 ppt
0.0 – 100.0 ppt
8
10
0.0 ppt
100.0 ppt
The 4-20 mA current loop can drive a load resistance of no more than 200 Ω.