Communication

3

30 Series Programming Instruction Manual

Data String Format Table

Prefix (3 bytes)

,

Data

,

Data

,

Data

,

P(S)(R) S = Sensor number R = Range number

,

8 bytes of direct ADC output

,

8 bytes of displayed data

,

8 bytes of internal offset/tare value

*

mA(R)

R = Range number

,

8 bytes of direct ADC output

,

8 bytes of displayed data

n/a 18 bytes of ASCII spaces

*

P(S)T S = Sensor number

,

8 bytes of direct ADC output of sensor
temperature signal

n/a 18 bytes of ASCII spaces

*

Amb

,

8 bytes of direct ADC output of sensor
temperature signal

n/a 18 bytes of ASCII spaces

*

BxC

x = Z for zero -or- x = S for span
C = ADC channel number

n/a

27 bytes of ASCII spaces.
This is a background calibration of the
analog to digital converter - no data.

*

,

= Delimiter

n/a = no delimiter is present for marked location in that string

*

= Terminator:

>

or

<

or

?

> = good battery

< = low battery

? = expired battery (readings are not valid)

Model Ranges

The standard Model 33 range serial numbers

2262

*

-xxxxxx are:

LP Sensor

HP Sensor

(Eight Ranges)

(Three Ranges)

(Five Ranges)

P11= inH

2

O

P21= bar

P21= kg/cm

2

P12= mbar

P22= kPa

P22= bar

P13= kg/cm

2

P23= PSI

P23= Mpa

P14= mmHg

P24= kPa

P15=mmH

2

O

P25= PSI

P16= kPa

P17= inHg

P18= PSI