Overview, Introduction – Crystal T-975-CPF Pneumatic Hand Pump User Manual

Page 3

Advertising
background image

Overview

1

T-900 Series Pump Operation Manual

Overview

„„„„

„

INTRODUCTION

The T-900 Series calibration hand pump, designed by Mansfield & Green, generates pressure for verifying, adjusting and calibrating mechanical and electronic
pressure measurement devices.

This hand pump is suitable for pressure tests in laboratory and field settings.

The T-900 Series calibration hand pump is easy to operate and allows for precise pressure generation. Combination models include a shuttle valve to allow
vacuum generation as well. CPF versions include

Crystal Pressure Fittings (CPF),

which allow users to produce leak-free seals without tools or thread tape.

CPF fittings also include a self-venting weep hole to help assure a safe disconnection from a pressurized system.

The T-900 Series calibration hand pump may be ordered as part of a Pump System, complete with a JOFRA or Crystal Pressure Indicator. T-900 Series Pump
Systems include the most commonly used pressure fittings, seals, etc. All packaged in a carrying case with custom insert.

X

T-900 Series Pumps and Pump Systems

Pressure

Vacuum

Pressure Range

Pump System

T-960

„

0 to 2 bar / 0 to 30psi

AXX

T-965

„

„

-0.85 to 2 bar / -25 inHg to 30 psi

BXX

T-970

„

0 to 40 bar / 0 to 580 psi

AHX

T-975

„

„

-0.91 to 40 bar / -27 inHg to 580 psi

BHX

T-975-CPF

„

„

-0.91 to 40 bar / -27 inHg to 580 psi

BHX and HOX

Advertising