TROUBLESHOOTING

Unstable Pressure

X

Problem:

Pressure cannot be generated correctly, or set pressure does not remain stable.

X

Solutions:

•

Check that all adapters and pressure fittings have been tightened sufficiently to eliminate leaks.

•

Confirm test fluid does not contain air (air in the fluid can increase adiabatic effects).

You may have to adjust the pressure a few times before the

pressure is stable enough for calibration. Replace test fluid if necessary.

•

Adiabatic and thermodynamic effects may be affecting the pressure reading.

High pressure hydraulic systems are sensitive to temperature and move-

ments. This effect will disappear after a period of time.

If the pressure is still not stabilizing, the system may be in need of service. Please contact your local distributor for advice.

Note:

Always readjust the pressure after 5 to 10 minutes to allow time for settling of above effects.

„„„„

„

MAINTENANCE

Users should/must carry out the following cleaning procedures when required:

Standard exterior cleaning . . . . . . . Clean using a soft, damp cloth.

Heavily soiled cleaning . . . . . . . . . . . The instrument may also be cleaned using isopropyl alcohol when heavily soiled.