10

Notes
1) „Over“ (data type DISCRETE) for invalid pressure value
2) For the Data Type key, refer to the table on Page 18.
3) User-interval period will, for example, be displayed as follows, “user 01:15:00”


3.6.2

Transfer of data to a PC

(with HM35 Communication Software)

1.

Install the IR (IrDA) -adapter according the instructions of the manufacturer.

2.

Install the HM35 Communication Software.

3.

Start the HM35 Communication Software.

4.

Place the instrument max. 20 cm from the IR (IrDA)-Adapter and switch it on. Ensure
a line-of sight connection between instrument and IR-adapter!
If there is no communication with the instrument for more than 2 minutes, the IR
interface of the instrument turns off automatically! By restarting the instrument the IR
interface is reactivated.



3.6.3 Deleting

data


1.

In

the

Menu,

select the Data Logging

→ Clear Memory function.

2.

Press the Clear key.




3.7 Communication

3.7.1 IR/RS232-Protocol

COM-Port Settings
Baudrate

9600

Data bits

8

Parity

no

Protocol

no

Stop bit

1


Communication Protocol

Coding
The characters are transfered as ASCII-Code.