5: Troubleshooting

35

5: Troubleshooting

5.1

Overview

If you have a problem with your flow monitoring system it can be due to any of the following:

Faulty instrument

Blank instrument display:

• Loss of power supply to the instrument.
• Internal power supply fuse ruptured.

Scrambled instrument display:

• Reboot the instrument by temporarily disconnecting its power supply.

Incorrect setup

A low, or zero, signal could be caused by incorrect set-up such as:

• Incorrect site data entered into the instrument.
• Incorrect or non-matching ultrasonic transducers selected for use.
• Incorrectly fitted transducers – lack of couplant applied, incorrect spacing, insecure attachment.
• Poor connections between the probes and the instrument.

Application problem

If you are certain that the instrument is healthy and suitably set-up for the current site; and the probes are
properly assembled and fitted correctly, there could be an application problem concerned with the site.

Check such conditions such as:

Poor pipe outer surface quality

• Uneven surface preventing good surface contact with the transducer.
• Flaking paint (should be removed).
• Variable air gap in concrete-covered pipes affecting the ultrasonic signal quality.

Poor internal pipe construction

• Rough internal pipe walls affecting fluid flow (see roughness factor).
• Internal welds positioned in the transducer signal path affecting the signal quality.
• The ‘drippings’ in galvanized-dipped pipes or other irregularities interfering with the signal path.

Incorrect probe location

• Transducers located too close to bends or valves, disturbing the flow profile.
• Transducers located too close to insertion probes, disturbing the flow profile.
• For horizontal pipework, transducers should not be positioned on the top of the pipe.

Poor fluid conditions within the pipe

• Fluid contains bubbles, high particle density or sludge.
• Air in the top of the pipe.

Low fluid flow within the pipe

• Pipe obstructions.
• Malfunctioning valve not opening fully (or closed inadvertently).

Liquid content problems

• Multiple liquid contents do not comply accurately to expected sound speed criteria.