Kaman KDM-8206 User Manual
Page 6

Kaman Precision Products
PART NO: 860516-001
www.kamansensors.com
Last Revised: 6/20/2011
6
All connections in the table are present in a Kaman supplied rack enclosure; no additional wiring must
be specified. Some comments on the connections in this table are as follows:
1. NC = Not Connected
2. Single channel voltage output is between pins A/C 14 and Ground. This is connected to the
upper BNC connector on rear panel of the rack.
3. Differential voltage output is between pins A/C 14 (V+Out) and A/C 16 (V-Out). Output at pins
A/C 16 is not present unless enabled on the Measuring Channel PCB. Refer to section 3.4 of
this manual. V-Out is connected to the lower BNC connector on the rear panel of the rack.
The lower BNC connector is not installed unless differential output is requested.
4. Demodulator output on pin A 24 can be connected to the BNC connector on the rear panel of
the rack in place of V+Out on request. Modification to the channel backplane in the rack is
required.
A good quality linear or switching supply with an output of +15 and -15 VDC is recommended for best
(lowest noise) performance.
Sensor installation recommendations and guidelines are given in Kaman Measuring System’s
Inductive Technology Handbook. A copy of this handbook is available from the website.