Part 2 – connections – Kaman digiVIT User Manual

Page 5

Advertising
background image

www.kamansensors.com

PART NO: 860525-001

Last Revised 9/16/2013

- 5 -

PART 2 – CONNECTIONS

The digiVIT I/O connections are all through a 10 pin removable terminal block. The DigiVIT
sensor is connected through an SMA coaxial connector on the opposite side of the enclosure.
An RJ45 Ethernet connection is provided for UDP/IP communication.


Pin Name

Function

1 +24V

+24V Input @ 0.13A
(must accommodate power-on surge current up to 300mA)

2 Gnd

Ground

3

Vout

Voltage Output (0-5, +/-5, 0-10, +/-10)

4 Gnd

Ground

5

Iout

Curent Loop Output (0-20mA, 4-20mA)

6 Gnd

Ground

7 NC

No

Connection

8

IO

IO Bit for special functions -- standard setup for Autozero when grounded.

9

Relay

Solid State Relay Connection

10

Relay

Solid State Relay Connection


Table 1 Terminal Block Pin-Out