Part 6 - integral power supply, Part 6 integral power supply, 1 power supply – Kaman KDM-8206 User Manual

Page 25

Advertising
background image

 

Kaman Precision Products

PART NO: 860515-001

www.kamansensors.com

Last Revised: 6/20/2011

25

PART 6 - INTEGRAL POWER SUPPLY

WARNING

: The integral power supply utilizes high voltage

(100 to 240VAC). Unplug the unit before servicing.

6.1 Power Supply

The power supply for the KDM-8206 rack enclosures is integral to the chassis and
mounted internally on the side panel. The KDM-8206 power supply provides a ±
15VDC regulated output. The same power supply is utilized for the Full 19” Sub-Rack,
the 1/2 Rack Instrument Enclosure, and the 3/4 Rack Instrument Enclosure. It has
capacity to power 1 to 12 Measuring Channels, or the Optional Display Module and 1 to
10 Measuring Channels. Power supply specifications are:

Input Power:

100VAC ±10% to 240VAC ±10%, 50/60 Hz, 76 VA

Output Power:

+15VDC ± 0.5V @ 1A (Peak Load 1.5A)

-15VDC ± 0.5V @ 0.67A (Peak Load 1A)

No modifications are required to operate the KDM-8206 Sub-Racks or Instrument
enclosures for input voltages between 100 and 240VAC.

DC power is distributed from the power supply via a 4 pin connector to the Channel 1
back plane PCB in the first slot. The back plane has a mating connector and can be
either a Measuring Channel back plane PCB or a Display Meter back plane PCB. It is
then distributed to all channels in the rack via interconnecting back plane PCB
connectors for the Measuring Channels.

The power supply is a unique design to provide specific voltages and currents required
for the Measuring Channels. It is physically configured to mount on the side rail of the
Instrument Enclosures and Sub-Rack.

For Figure 21 below, Power Supply connections labeled are:

J1 DC output to Back Plane PCB of Measuring Channel No. 1
G1 Input ground from Power Entry Module
G2 Chassis ground
G3 Chassis ground
L Line AC input from Power Entry Module
N Neutral AC input from Power Entry Module

Advertising