Product overview, Product user guide – MadgeTech RFRTDTemp2000A User Manual

Page 4

Advertising
background image

Product User Guide

Product Overview

The RFRTDTemp2000A is an RTD-based, wireless temperature data logger with display.
This data logger is ideal for applications that require high accuracy temperature monitoring
such as within laboratories, pharmaceutical and biotech industries. The RFRTDTemp2000A
accepts 2, 3 or 4 wire 100 Ω platinum RTD’s and features a device accuracy of +/-0.05 °C.

Selection Buttons

The logger is designed with 3 direct selection buttons: Scroll, Units and Wireless.
Scroll: Allows user to scroll through current readings, average stats and device status

information displayed on the LCD Screen.

Units: Allows user to change displayed units of measure to either °F or °C.
Wireless: Push and hold this button for 5 seconds to activate or deactivate

wireless communication.

Manual Reset Option: Allows user to manually reset the statistics within the device to 0

without needing to use the MadgeTech software. Any data recorded up to that point is
recorded and saved. To apply the manual reset, press and hold the scroll key down for
three seconds.

LED Indicators

Status: Green LED blinks every 5 seconds to indicate unit is logging.
Wireless: Blue LED blinks every 15 seconds to indicate unit is operating in wireless mode.
Alarm: Red LED blinks every 1 second to indicate an alarm condition.

2

Advertising