Unit

HygroPalm HP22-A hand-held indicator:

User Guide

Instruction Manual

Document Type

Page 10 of 16

Document title


© 2010-2012; Rotronic AG E-M-HP22-V2_13

4.9 Internal menu

Note: Unauthorized access to the menu can be prevented by disabling the “display menu” setting
(use the HW4 software > Device Manager > Display)

Main Menu

Menu Items

Selections / Information

Notes

Device Settings

Units

Metric / English

Back Light

Key Press / On / Off

Display backlight mode

Contrast

LC display contrast adjustment

Trend

On / Off

Trend indication on the display

AutoPwrOff

OFF
1 min
5 min
15 min
30 min
45 min
60 min

The HP22 powers itself down after
a user specified time period

Device Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Type

Device type

Name

Device name

User defined

Probe Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Name

Device name

User defined

SensorTest

Humidity sensor status

Off / Good / SQ-Tuned / Bad

Record

On / Off

Data recording by the probe
(max. 2000 values)

Humidity Adjust

RefValue

Humidity reference value

± 0.1 %RH steps

Acquired

Number of cal. points in probe
memory

<Acquire>

Save cal. point to probe memory

<Delete>

Erases all calibration points

<Adjust>

Effect depends on number of
calibration points

Temperature Adjust

RefValue

Temperature reference value

± 0.1 ˚C steps

<Adjust>

1-point adjustment only (offset)


o

Record: both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software