MEASUREMENT LOOP VALIDATION

Functions

Description

► Simulator mode

Used to make the HygroMet generate fixed values for the
humidity, temperature and calculated parameter. Can be
configured, enabled or disabled

DEVICE SAFEGUARDS

Functions

Description

► Device write protection

Used to protect the HygroMet with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled

PROCESS PROTECTION AND PROTECTION OF OTHER DEVICES

AirChip 3000 Functions

Description

► Limit humidity output to 100 %RH

This probe function is used to prevent the humidity signal from
exceeding 100 %RH when condensation forms on the sensor.
Can be enabled or disabled

► Out-of-limit value alarm

Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled

Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display

► Bad sensor alarm

This is a built-in probe function. Cannot be disabled

A bad humidity or temperature sensor triggers a digital alarm,
provided that the HygroMet is set to monitor probe alarms

► Fail safe mode

Used to specify a "safe" fixed value for humidity and for
temperature (HygroMet or probe) in the event of:

o

Loss of communication with the probe (HygroMet function)

o

Sensor failure (probe function)


Can be configured, enabled or disabled

4.3 Factory default settings


Configurable Settings

Applicability

Factory default

Unit system (Metric or English)

2-wire, 3-wire

As per ordering code

Analog signal type (4…20 mA or other)

3-wire

As per ordering code

Psychrometric calculation

2-wire, 3-wire

As per ordering code

Fixed pressure value

2-wire, 3-wire

1013.25 hPa or 29.92 In Hg

Output 1 parameter, scale and unit

2-wire, 3-wire

As per ordering code (%RH or other)

Output 2 parameter, scale and unit

2-wire, 3-wire

Temperature, unit as per ordering code

Communication protocol

3-wire

RO-ASCII

RS-485 address

3-wire

0

Device name

2-wire, 3-wire

Instrument model