EA003 OPTICAL TACHOMETER CALIBRATION ADAPTER

OPERATION MANUAL

Transmille Ltd.

Page 7


The main screen comprises of a main reading display, range selection buttons and
function selection buttons.

Sampling is indicated on the main

display by

Functions : Current Measurement


The current measurement function provides five ranges to measure current:

100uA • 10uA • 1uA • 100nA • 10nA

Functions: Resistance Measurement


The resistance measurement function provides five ranges. On initial selection of
resistance, the program will indicate a voltage should be set and the range buttons will
display ‘No Volts’.

Selecting a test voltage using the voltage selection controls :


Select a voltage from the drop down list, or by entering a voltage using
the keyboard.

Valid voltages are from 20V to 1000V.



Once a voltage is selected the range buttons will display the resistance ranges, for