S p e c i f i c a t i o n s

Sensor Reconditioning
Over time, the internal sensor may be compromised as a result of pollu-
tants, chemical vapors, and other environmental conditions which can
lead to inaccurate readings. To recondition the internal sensor, please
follow the procedure below:

Dry the logger at 176˚F (80˚C) at <5% RH for 36 hours followed by
70˚ - 90˚F (20˚ - 30˚C) at >74% RH for 48 hours (for re-hydration).

If permanent damage to the internal sensor is suspected, replace the
Logger immediately to ensure accurate readings.

THL1-MAN

P. 3

Relative Humidity

Overall range

0 to 100%

Accuracy (0 to 20 and 80 to 100%)

±5.0%

Accuracy (20 to 40 and 60 to 80%)

±3.5%

Accuracy (40 to 60%)

3.0%

Temperature

Overall range

-40˚ to 158˚F
(-40˚ to 70˚C)

Accuracy (-40˚ to +14 and 104˚ to 158˚F)

±3.6˚F

Accuracy (+14˚ to +104˚F)

±1.8˚F

Accuracy (-40˚ to -10˚ and +40˚ to +70˚C)

±2˚C

Accuracy (-10˚ to +40˚C)

±1˚C

Dew Point

Overall range

-40˚ to 158˚F

Temperature

(-40˚ to 70˚C)

Accuracy (25˚C, 40 to 100% RH)

±4.0˚F (2.0˚C)

Logging Rate

Selectable sampling interval: From 2 seconds up to
24 hours.

Operating

-31˚ to 175˚F (-35˚ to 80˚C)

Temperature
Battery Type

306V lithium (1/2AA) (SAFT LS14250, Tadiran TL-5101
or equivalent)

Battery Life

1 year (typ.) depending on logging rate, ambient
temperature and use of alarm LEDs.

Dimensions

101 x 25 x 23 mm (4 x 1 x 9”)

Weight

172 g (6 oz)