Introduction – Agilent Technologies E4400-60597 User Manual
Page 3

Agilent Technologies Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators
Check Signal Generator Functionality
3
Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal
Generators
Introduction
The instructions in this firmware upgrade guide enable you to install the PSG/ESG Upgrade Assistant 5.12 on your
computer from either the Agilent Technologies website or a CD-ROM and upgrade firmware on ESG-B signal generators
with <B.03.20.
NOTE
For Agilent E44xxB ESG Signal Generators Models Only
When upgrading Agilent E44xxB ESG signal generators from firmware versions <B.03.20, see,
Firmware Installation For Versions <B.03.20” on page 14
To upgrade firmware on PSG/ESG signal generators, perform the following procedures:
1. Check signal generator functionality on
.
2. Confirm required equipment on
.
3. Connect an RS-232, GPIB, or LAN interface to a PC on
.
4. Install PSG/ESG Upgrade Assistant 5.12 and firmware files on
.
5. Use the PSG/ESG Upgrade Assistant 5.12 with firmware files on
.
6. Check signal generator functionality on
.
Product Affected: . . . . . . . . . . . . . . . . . . . . . . . . .
All
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . .
All
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
All except Option H98
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . .
(X) Agilent Technologies Service Center
(X) Personnel Qualified by Agilent Technologies
(X) Customer
Estimated Installation Time: . . . . . . . . . . . . . . .
75 minutes (RS-232), 30 Minutes (GPIB), 3 minutes
(LAN)
Estimated Verification Time: . . . . . . . . . . . . . . .
1 minute