2. function list – A&D FS-D User Manual
Page 13

11
6-2. Function list
Item
Function number
Description
C01- 0
CAL switch enabled,
0
key enabled
C01- 1
CAL switch disabled,
0
key disabled
Calibration
C01- 2
CAL switch enabled,
0
key disabled
Do not set
“C01-1”
C02- 0
6 kg (15 lb)
C02- 1
15 kg (35 lb)
C02- 2
30 kg (70 lb)
C02- 3
60 kg (150 lb)
C02- 4
150 kg (350 lb)
C02- 5
300 kg (700 lb)
C02- 6
600 kg (1500 lb)
Capacity
C02- 7
1500 kg (3500 lb)
C03- 0
kg, g
Weighing display
unit
C03- 1
kg, g, lb, oz, lb-oz
See “11-1.
Capacity ...”.
C04- 0
±10% of the capacity
OIML R76 / NTEP
Power-on-zero
range
C04- 1
±50% of the capacity
Internal use
C05- 0
Do not change.
C06- 0
.
(point)
Decimal
separator
C06- 1
,
(comma)
C07- 0
ZERO does not clear tare weight
Tare clear by
ZERO operation C07- 1
ZERO clears tare weight
OIML R76
C08- 0
Tare weight is rounded to the weight
display count
NTEP
Tare operation
(Net weight
calculation)
C08- 1
Internal count is used for TARE weight.
OIML R76
C09- 0
Function disabled
Preset tare
function
C09- 1
Function enabled
C10- 0
lb-oz disabled
lb-oz unit
C10- 1
lb-oz enabled
C11- 0
±2% of the capacity
ZERO operation
range
C11- 1
Up to the capacity
C12- 0
“f02” setting specifies the resolution.
C12- 1
Fixed to “f02-0” (1/3,000~1/3,500)
C12- 2
Fixed to “f02-1” (1/6,000~1/7,500)
Display resolution
C12- 3
Fixed to “f02-2” (1/12,000~1/15,000)
C13- 0
Function disabled
Auto-tare function
C13- 1
Function enabled
C13-0 deletes
“f21” ~ “f23”.
Factory setting