10. Cut the catheter off of the Faller trocar at the end of the
barbed tip of the trocar.
CAUTIONS:
• Do not pull the catheter off.
• Do not cut the suture off.
• attempting to cut the suture off creates a very high risk
of damaging the catheter.
11. Check that the catheter is not twisted or kinked.
12. Do not suture the exit-site.
13. attach the appropriate connector to the catheter as
recommended by the catheter manufacturer. attach a cap
or transfer set to the connector.
14. Complete the balance of the catheter implantation
procedure as per established protocol or catheter
instructions for use.