Merit Medical Flex-Neck Classic ARC Catheter User Manual

Page 4

Advertising
background image

CAThETER CONNECTOR INSTRUCTION

A plastic connector is included with each Flex-Neck Perito-

neal Dialysis Catheter. A two-piece titanium connector for

Flex-Neck Adult, Adolescent, and Pediatric Peritoneal Dialy-

sis Catheters is available separately from Merit Medical.

After successful implantation of the Flex-Neck Peritoneal

Dialysis Catheter, attach a Merit Peritoneal Dialysis (PD)

catheter connector to the catheter.

Each catheter kit contains one Connector and one Cap.

• Tapered Tip (A)

A

• Raised Shoulder Ridge (B)

• Finger Grip (C)

C

B

• Threaded Luer (D)

D

• Cap (E)

E

1. Wet the Tapered Tip (A) of the Connector with sterile

saline or sterile water, and insert it into the catheter.

• Do not use any other lubricant.

• Do not use a twisting motion to force the catheter onto

the Connector.

• Push the Connector into the catheter with a single

forward motion.

2. Advance the catheter completely to the Raised Shoul-

der Ridge of the Connector. The catheter tubing must

completely pass over the Tapered Tip and to the Raised

Shoulder Ridge, but not beyond that onto the Finger Grip.

See diagram.

3. Pull carefully on the Connector and catheter to test the

strength of the connection.

4. Attach either the Cap (E), or a dialysis transfer set, to the

threaded Luer (D).

Catheter Cleaning and Care

All Flex-Neck Peritoneal Dialysis Catheters are made of

silicone. Exit-site cleaning agents that are compatible with

silicone catheters therefore may be acceptable for use

on Flex-Neck Peritoneal Dialysis Catheters. Such cleaning

agents include:

• Electrolytically-produced sodium hypochlorite solutions

(i.e., ExSept Plus®)

• Normal (sterile) saline

Cleaning agents that are non-irritating, non-toxic, anti-bac-

terial, and in liquid form are generally recommended.

The following cleaning agents are not compatible with

silicone catheters, and are not recommended for use with

Flex-Neck Peritoneal Dialysis catheters:

• Acetone or acetone-based products

• Povidone-iodine or iodine-based products

Merit Medical Systems, does not provide specific recom-

mendations or protocols for exit-site care and cleaning,

whether by the healthcare professional or by the patient.

Appropriate exit-site and catheter care treatment protocols

should be individualized for each patient, and established

by the patient’s physician(s), nurse(s), dialysis center(s),

and/or other relevant dialysis healthcare professionals.

ExSept Plus is a registered trademark of Alcavis HDC, LLC or

one of its affiliates.

Copyright © Merit Medical Systems, Inc. All rights reserved.

Manufacturer:

Merit Medical Systems, Inc.

1600 West Merit Parkway, South Jordan, Utah 84095 U.S.A.

1-801-253-1600 U.S.A. Customer Service 1-800-356-3748

Authorized Representative:

Merit Medical Ireland Ltd,

Parkmore Business Park West, Galway, Ireland

www.merit.com

ID 062613 402896001/A

Advertising