Alarm Unit Mounting Instructions -

Saddle Mount

1. Choose a secure location out of the patient’s

reach, preferably in the rear, below the chair seat

(Fig. 1).

2. Wrap the saddle strap firmly around a bar of

wheelchair or bed frame. Check that saddle is

secure and will not move.

3. Press hook material on alarm unit firmly to loop

material on saddle (Fig. 2).

4. Check that alarm unit and sensor cable are clear

of all moving parts of the chair.

Fig. 1 Wheelchair Mount

attached to lower, rear

wheelchair frame

Hook and Loop Mount

1. Choose site on head or footboard or other

mounting surface, i.e. nightstand, out of the

patient’s reach.

2. Clean site with alcohol wipe and let it dry.

3. Attach self-adhesive loop (fuzzy) patch (7). Peel

off backing and press into place on the mounting

surface.

NOTE: Adhesive on loop will not adhere

to vinyl or other slick surface.

4. Press the hook (rough) patch (8) on back of alarm

unit firmly to the loop patch on mounting surface.

Be sure alarm unit is secure and will not move.

Before Leaving Patient Unattended

ALWAYS follow these steps each time before leaving

patient unattended:

1. Connect a Posey sensor to the alarm unit. Check

to make sure that the sensor cable and RJ11 are

intact and undamaged.

2. Place weight on sensor pad.

3. Turn alarm unit ON and verify green light blinks.

4. Verify batteries are fresh. If you hear an

intermittent “CHIRP”, the battery is low and needs

to be changed.

5. Check to make sure the audible alarm is functioning

properly by applying and lifting weight off sensor in

several spots to activate the alarm.

6. DO NOT use the Posey KeepSafe if the audible

alarm does not sound.

7. Check to make sure sensor cable is out of the way

and does not pose a tripping hazard. If cable must

cross footpath, use duct tape or similar product

to secure cable to floor and reduce hazard.

NOTE: If the sensor is damaged or disconnected

while in use, alarm will sound to notify caregiver.

Cleaning Instructions

• Alarm Unit and Saddle Mount – Wipe clean with

disinfectant liquids that DO NOT contain Phenol or

Benzyl. DO NOT submerge in liquid or sterilize with heat.

• Hook and Loop – Use stiff brush to remove dust and lint.

Fig. 2

Limited Lifetime Warranty:

The Posey Company warrants to the original

purchaser, the alarm unit to be free from defects

in materials and workmanship for the life of the

unit. Warranty void if unit is opened, damaged, or

submerged. This warranty gives you specific legal

rights, and you may also have other rights, which vary

from state to state.

• Read sensor and alarm instructions prior to use.

NEVER connect Posey sensors to other

manufacturers’ alarm units.

NEVER connect other manufacturers’ sensors to

Posey alarm units.

ALWAYS test system per instructions prior to use.

NEVER pull the sensor cable from the alarm. Doing

so will damage the connections and may prevent

the alarm from sounding.

ALWAYS use the lever to

release the RJ11 from the alarm unit.

• This device may not be suitable for use with all

“high fall risk” patients.

• This device should not be a substitute for

maintaining the routine visual monitoring protocol

by healthcare personnel.

• This device provides the earliest possible warning

that a patient is at immediate risk of falling. This

device does not prevent falls.

• When used, be sure restraint straps DO NOT cross

over sensor pad! Restraint straps may prevent alarm

activation if patient falls out of bed or chair.

• The alarm may fail to sound if the sensor pad is

folded or bent.

ALWAYS make sure the sensor cable does not

contact moving parts of the chair.

ALWAYS make sure the sensor cable is clear of all

moving parts of the bed to prevent damage to the

sensor and cable when the bed is articulated.

ALWAYS make sure the sensor cable does not pose

a tripping hazard.

Battery Replacement

The alarm unit will emit an audible “CHIRP”, indicating a

low battery. To replace the battery:
1. Turn the alarm OFF.

2. Press down on arrow and slide the battery door open.

3. Remove the battery from the compartment and hold

in one hand with thumb on the terminal cap. Battery

connection wires should be away from the thumb.

DO NOT pull the terminal cap wires.

4. Use thumb to push the

terminal cap up, off the

battery (Fig. 3).

5. Connect terminal cap to the

new battery. Make sure the

battery is securely connected

to the terminal cap and insert

into alarm unit.

6. Slide battery door closed.

7. Test system to verify it is working properly before

leaving patient unattended.

Batteries can explode or leak and cause damage to

alarm or skin if installed backwards, disassembled,

fully discharged or exposed to water, fire or high

temperatures.

• ALWAYS install a completely new set of batteries

when the chirping due to low battery power sound

begins; DO NOT replace a single cell, but all cells in

the alarm unit.

Installing batteries incorrectly can cause batteries to

utilize more power than normal and discharge sooner,

causing possible corrosion. Take care when installing

new batteries.

Fig. 3

Bed Pad Sensor Instructions

1. Place the sensor pad on the mattress directly

under the patient’s weight. Most patient weight is

under the buttocks. See sensor pad label for other

positions when head of bed is elevated.

2. Use sensor pad clips at each end to attach sensor

to the mattress.

3. Be sure to route sensor cable to the alarm unit so

it is clear of all moving parts of bed to prevent

damage to the cable or the pad when bed is raised

or lowered, or when head or knee sections are

operated.

4. Alarm may fail to sound if the sensor pad is bent

or folded under the patient, or if there is weight

from an object on the pad.

5. Make sure sensor cable is not in the footpath and

does not pose a tripping hazard.

Proper Handling And Use

If the Posey KeepSafe is subjected to severe

mechanical shock, such as dropping, or is submerged

in liquid, it may stop functioning as designed. Visually

inspect the unit for cracks in the case, missing

screws, missing battery door, broken wires or signs

of exposure to liquid. After each incident, you must

verify that the unit is working properly.

Storage Instructions

• This device is designed for use in normal indoor

environments.

• This device may be stored in ambient warehouse

temperatures at normal humidity levels. Avoid

excess moisture or high humidity that may damage

product materials.

Disposal

Dispose of per your facility policy or applicable local

code for BIOHAZARDOUS material.