Kestrel

®

4200 Pocket Air Flow

®

Tracker

21

default setting: Metric, Imperial or Defaults. Press the or button to reset the factory setting. Press the

button to return to the Main Setup Menu.

APPLICATION EXAMPLES

This section provides examples of applications where a Kestrel Meter might be used, and the appropriate

memory settings.
HVAC - Environmental Monitoring

Auto Store

On

Store Rate

5 min

Overwrite

On

Man Store

Off

These settings will record conditions every five minutes, for a total storage of almost 7 days. You can

monitor the conditions in a laboratory or manufacturing plant, both day and night, to determine if the

climate control is working properly. Or you can examine the effect on the environment when employees

enter and exit the building.

HVAC/R - System Balancing

Auto Store

Off

Store Rate

Overwrite

Off

Man Store

On

These settings will require you to press the Manual Store Button in order to store any data at a duct, hood,

vent, or other system location. The meter will not store any data automatically. Be sure to record the

location and date/time of storage for reference when reviewing the data. After storing the conditions at

each location, simply review the data and balance the system.

MEMORY CAPABILITIES

Store Rate

Total Memory

2 sec

53 min, 20 sec

5 sec

2 hr, 13 min, 20 sec

10 sec

4 hr, 26 min, 40 sec

20 sec

8 hr, 53 min, 20 sec

30 sec

13 hr, 20 min

1 min

1 day, 2 hr, 40 min

2 min

2 days, 5 hr, 20 min

5 min

5 days, 13 hr, 20 min

10 min

11 days, 2 hr, 40 min