Post calibration actions 6-10, Rf power/alc calibration adjustments – Anritsu 373XXA User Manual
Page 90

quence: H/W CALIBRATIONS, and then SOURCE ALC
CALIBRATION (below).
Step 3.
Follow the directions displayed on the screen to set-up and
connect the power meter to the 373XXA. Refer also to Fig-
ure 6-2. Select START ALC CALIBRATION from menu.
Step 4.
Follow the directions displayed on the 373XXA CRT screen,
until calibration is completed.
Step 5.
If the software allows a Port 2 ALC Calibration, perform it
now.
Post Calibration
Actions
After the calibration process is completed, perform the following ac-
tions, as appropriate:
RF POWER/ALC CALIBRATION
ADJUSTMENTS
6-10
373XXA MM
7
8
9
4
5
6
1
2
3
0
.
-
SC6230 Sensor
ANRITSU 373XXA
VNA
Dedicated GPIB
ANRITSU ML24XXA
Power Meter
Figure 6-2.
Equipment Set-Up for RF Power/ALC Calibration
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE SOURCE
CONTROL
PRESS <ENTER> TO
SELECT
DIAGNOSTICS
START SELF TEST
READ SERVICE LOG
INSTALLED OPTIONS
PERIPHERAL TESTS
TROUBLESHOOTING
(FOR SERVICE
ONLY)
H/W CALIBRATIONS
(FOR SERVICE
ONLY)
PRESS <ENTER> TO
SELECT
HARDWARE
CALIBRATIONS
(FOR SERVICE
ONLY)
LO1
CALIBRATION
LO2
CALIBRATION
SOURCE FREQUENCY
CALIBRATION
SOURCE ALC
CALIBRATION
DISK OPERATIONS
PREVIOUS MENU
PRESS <ENTER> TO
SELECT
HARDWARE
CALIBRATIONS
(FOR SERVICE
ONLY)
START SOURCE
FREQ CALIBRATION
PREVIOUS MENU
PRESS <ENTER> TO
SELECT
Set-Up