Post calibration actions 6-10, Rf power/alc calibration adjustments – Anritsu 373XXA User Manual

Page 90

Advertising
background image

quence: H/W CALIBRATIONS, and then SOURCE ALC
CALIBRATION
(below).

Step 3.

Follow the directions displayed on the screen to set-up and
connect the power meter to the 373XXA. Refer also to Fig-
ure 6-2. Select START ALC CALIBRATION from menu.

Step 4.

Follow the directions displayed on the 373XXA CRT screen,
until calibration is completed.

Step 5.

If the software allows a Port 2 ALC Calibration, perform it
now.

Post Calibration

Actions

After the calibration process is completed, perform the following ac-
tions, as appropriate:

RF POWER/ALC CALIBRATION

ADJUSTMENTS

6-10

373XXA MM

7

8

9

4

5

6

1

2

3

0

.

-

SC6230 Sensor

ANRITSU 373XXA

VNA

Dedicated GPIB

ANRITSU ML24XXA

Power Meter

Figure 6-2.

Equipment Set-Up for RF Power/ALC Calibration

OPTIONS

TRIGGERS

REAR PANEL

OUTPUT

DIAGNOSTICS

MULTIPLE SOURCE

CONTROL

PRESS <ENTER> TO

SELECT

DIAGNOSTICS

START SELF TEST

READ SERVICE LOG

INSTALLED OPTIONS

PERIPHERAL TESTS

TROUBLESHOOTING

(FOR SERVICE

ONLY)

H/W CALIBRATIONS

(FOR SERVICE

ONLY)

PRESS <ENTER> TO

SELECT

HARDWARE

CALIBRATIONS

(FOR SERVICE

ONLY)

LO1

CALIBRATION

LO2

CALIBRATION

SOURCE FREQUENCY

CALIBRATION

SOURCE ALC

CALIBRATION

DISK OPERATIONS

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

HARDWARE

CALIBRATIONS

(FOR SERVICE

ONLY)

START SOURCE

FREQ CALIBRATION

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

Set-Up