Aalborg GFM User Manual

Page 10

Advertising
background image

4.

SPECIFICATIONS

FLOW MEDIUM: Please note that GFM17/37/47/57/67/77 Mass Flow Meters are designed to
work with clean gases only. Never try to meter or control flow rates of liquids with any GFM's.

CALIBRATIONS: Performed at standard conditions [14.7 psia (1.01 bars) and 70

F

F

(21.1

F

C)] unless otherwise requested or stated.

ENVIRONMENTAL (per IEC 664): Installation Level II; Pollution Degree II.

ACCURACY: ±1.5% F.S. (optional enhanced accuracy ±1% F.S.).

REPEATABILITY: ±0.25% of full scale.

TEMPERATURE COEFFICIENT: 0.15% of full scale/

F

C.

PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar).

RESPONSE TIME: 800 ms time constant; approximately 2 seconds to within ±2% of set
flow rate for 25% to 100% of full scale flow rate.

GAS PRESSURE: 1000 psig (69 bars) GFM 17, 37 and 47; 500 psig (34.5 bars) GFM 57,
67 and 77; optimum pressure is 20 psig (1.4 bars).

GAS AND AMBIENT TEMPERATURE: 32

F

F to 122

F

F (0

F

C to 50

F

C).

14

F

F to 122

F

F (-10

F

C to 50

F

C) - Dry gases only.

RELATIVE GAS HUMIDITY: Up to 70%.

LEAK INTEGRITY: 1 x 10

- 7

sccs He max. to the outside environment.

ATTITUDE SENSITIVITY: Incremental deviation of up to 1% full scale from stated accura-
cy, after re-zeroing.

OUTPUT SIGNALS: Linear 0-5 VDC (1000

Ω minimum load impedance) and 4-20 mA

self-powered sourcing type, non-isolated (0-500 loop

Ω resistance); 20 mV peak to peak

max noise.

Contact your distributor or Aalborg

7

for optional RS232 or RS-485 interfaces.

6

TRANSDUCER INPUT POWER:

GFM models with 12 VDC power input (serial numbers below 340621-1):

12 VDC, 200 mA maximum;

GFM models with 24 VDC power input (serial numbers below 340621-1):

+24 VDC, 200 mA maximum;

GFM models with universal power input (serial numbers above 340621-1):

any voltage between +12 and +26 VDC, 200 mA maximum;