TABLE VI AFC SOLENOID VALVE ORIFICE SELECTION TABLE

8.

TROUBLESHOOTING

8.1

Common Conditions

Your Mass Flow Meter/Controller was thoroughly checked at numerous quality
control points during and after manufacturing and assembly operations. It was cal-
ibrated in accordance to your desired flow and pressure conditions for a given gas
or a mixture of gases.

It was carefully packed to prevent damage during shipment. Should you feel that
the instrument is not functioning properly please check for the following common
conditions first:

✓

Are all cables connected correctly?

✓

Are there any leaks in the installation?

✓

Is the power supply correctly selected according to requirements?
When several meters are used a power supply with appropriate
current rating should be selected.

✓

Were the connector pinouts matched properly?
When interchanging with other manufacturers' equipment, cables
and connectors must be carefully wired for correct pin configurations.

✓

Is the pressure differential across the instrument sufficient?

20

ORIFICE PART NUMBER

FLOW RATE [N2]

OR.010

under 10 sccm

OR.020

10 to 1000 sccm

OR.040

1 to 5 slpm

OR.055

5 to 10 slpm

OR.063

10 to 15 slpm

OR.073

15 to 20 slpm

OR.094

20 to 50 slpm

OR.125

50 to 100 slpm