Use, Maintenance, and Care of the Devices

Cleaning Connectors

Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any
dirt or stubborn contaminants on a connector that cannot be removed with compressed
air or nitrogen. Refer to

Table 6-3 on page 6-5

for part numbers for isopropyl alcohol and

cleaning swabs.

a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.

b. Clean the connector threads.

c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,

low-pressure compressed air or nitrogen. Always completely dry a connector before
you reassemble or use it.

3. Clean the Mating Plane Surfaces

a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.

b. Clean the center and outer conductor mating plane surfaces. Refer to

Figure 2-1 on

page 2-3

. When cleaning a female connector, avoid snagging the swab on the center

conductor contact fingers by using short strokes.

c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,

low-pressure compressed air or nitrogen. Always completely dry a connector before
you reassemble or use it.

4. Reinspect

Inspect the connector again to make sure that no particles or residue are present.