Programming, Maintenance and calibration – Campbell Scientific 108-LC Temperature Probe for MetData1 User Manual
Page 9

Model 108-LC Temperature Probe for MetData1
5
5. Programming
Information in this section is not necessary when programming
the MetData1 with the Short Cut Program Builder software.
Instruction 5 (AC Half Bridge) is used to measure the 108 temperature probe.
Instruction 55 (polynomial) is used to find the temperature in degrees Celsius.
MetData1 Datalogger Program Example
01: AC Half Bridge (P5)
1:
1
Reps
2:
3
± 25 mV Slow Range
3:
8 or 9**
In Chan
4:
3
Ex Chan Option
5:
1000
mV Excitation
6:
11*
Loc [ Tmp108C ]
7:
200
Mult
8:
0
Offset
02: Polynomial (P55)
1:
1
Reps
2:
11*
X Loc [ Tmp108C ]
3:
11*
F(X) Loc [ Tmp108C ]
4:
-26.97
C0
5:
69.635
C1
6:
-40.66
C2
7:
16.573
C3
8:
-3.455
C4
9:
.301
C5
* Proper entries will vary with input location assignments.
** Specify input channel #8 for connector #4. Specify input channel #9 for connector #7.
TABLE 5-1. Polynomial Coefficients
Coefficient
Value
C0
-26.97
C1
69.635
C2
-40.66
C3
16.573
C4
-3.455
C5
0.301
6. Maintenance and Calibration
The 108 Probe requires minimal maintenance. Check monthly to make sure the
radiation shield is free from debris.