Campbell Scientific 108-L Temperature Probe User Manual
Page 24

Appendix B. Example Programs
;{CR10X}
;Program measures one 108 temperature probe once a second
;and stores the average temperature every 60 minutes.
;Wiring Diagram
;==============
;108 Probe
;
; Wire
; Color
Function
CR10X
; -----
--------
-----
; Black
Voltage-excitation input E1
; Red
Analog-voltage output
SE1
; Purple Bridge resistor lead
AG
; Clear
EMF Shield
G
*Table 1 Program
01: 1.0000 Execution Interval (seconds)
1: AC Half Bridge (P5)
1: 1
Reps
2: 23
25 mV 60 Hz Rejection Range
3: 1
SE Channel
4: 1
Excite all reps w/Exchan 1
5: 1000
mV Excitation
6: 1
Loc [ T108_C ]
7: 200
Multiplier
8: 0
Offset
2: Polynomial (P55)
1: 1
Reps
2: 1
X Loc [ T108_C ]
3: 1
F(X) Loc [ T108_C ]
4: -26.97 C0
5: 69.635 C1
6: -40.66 C2
7: 16.573 C3
8: -3.455 C4
9: 0.301
C5
3: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
4: Set Active Storage Area (P80)
1: 1
Final Storage Area 1
2: 101
Array ID
5: Real Time (P77)
1: 1220
Year,Day,Hour/Minute (midnight = 2400)
6: Average (P71)
1: 1
Reps
2: 1
Loc [ T108_C ]
B-2