7. wiring for cr10x example program – Campbell Scientific 227-L Delmhorst Cylindrical Soil Moisture Block User Manual
Page 20

227 Delmhorst Cylindrical Soil Moisture Block
TABLE 7-7. Wiring for CR10X Example Program
Color Function
CR10(X)
Black Excitation
E1
Red Signal
SE1
White Signal
Ground
AG
Clear Shield
G
*Table 1 Program
01: 10.0000 Execution
Interval
(seconds)
01: AC Half Bridge (P5)
;Measure and store Vs/Vx
1: 1
Reps
2: 14
250 mV Fast Range
3: 1
SE
Channel
4: 1
Excite
all
reps
w/Exchan
1
5: 250
mV
Excitation
6: 1
Loc [ Rs ]
7: 1
Mult
8: 0
Offset
02: BR Transform Rf[X/(1-X)] (P59)
;Convert Vs/Vx to Rs
1: 1
Reps
2: 1
Loc [ Rs ]
3: 1
Multiplier
(Rf)
03: If (X<=>F) (P89)
;If Rs < 17, Use 10 bar polynomial
1: 1
X Loc [ Rs ]
2: 4
<
3: 17
F
4: 30
Then
Do
04: Z=X*F (P37)
;Scale Rs for polynomial
1: 1
X Loc [ Rs ]
2: .1
F
3: 2
Z Loc [ WatPoten ]
05: Polynomial (P55)
;Convert Rs to bars with 10 bar polynomial
1: 1
Reps
2: 2
X Loc [ WatPoten ]
3: 2
F(X) Loc [ WatPoten ]
4: .15836 C0
5: 6.1445 C1
6: -8.4198 C2
7: 9.2493 C3
8: -3.1685 C4
9: .33392 C5
06: Else (P94)
;If Rs > 17 load overrange value for potential
14