4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules
5.1 Edlog
Dataloggers that use Edlog include CR510, CR10(X), 21X, and CR7. The
Edlog instruction that measures strain gages is Instruction 6 – Full Bridge.
The Input Locations assignments used in CR10(X), 21X, and CR7 Examples
are listed in Table 5-1.
TABLE 5-1. Input
Locations Used in
CR10(X), 21X, and CR7
Examples
Addr Name
1 mVperV
2 mVperV_0
3 Vr
4 uStrain
5 Count
6 GF
7 _4e6
8 Mult
9 1_2Vr
10 Vr_1_2Vr
5.1.1 CR10(X)
;{CR10X}
;
*Table 1 Program
01: 1
Execution
Interval
(seconds)
;Other measurements could be inserted here or before the Output section
1: If Flag/Port (P91)
;On the first execution (Flag 1 is low)
1: 21
Do if Flag 1 is Low
;or when user sets Flag 1 low
2: 1
Call
Subroutine
1
;call the zeroing subroutine
2: Full Bridge (P6)
;Measure the strain gage
1: 1
Reps
2: 22
± 7.5 mV 60 Hz Rejection Range
3: 1
DIFF
Channel
4: 1
Excite all reps w/Exchan 1
5: 2500
mV
Excitation
6: 1
Loc [ mVperV ]
7: 1
Mult
8: 0
Offset
4