Campbell Scientific 43347 RTD Temperature Probe and 43502 Aspirated Radiation Shield User Manual

Page 37

Advertising
background image

Appendix D. Measure Two 43347-IX Probes Using One Current Excitation Channel

TABLE D-1. Wiring for Two 43347-IX Probes Example

Color Function

CR3000,

CR5000

Probe

#1

Red

Switched Current Excitation IX1

White Differential

High

1H

Green Differential

Low

1L

Black Excitation Return

Red of Probe #2

Clear Shield

Probe

#2

Red

Switched Current Excitation Black of Probe #1

White Differential

High

2H

Green Differential

Low

2L

Black Excitation

Return

IXR

Clear Shield

(2) 43502 Shields

White Tachometer

C1 for first probe, C2 for second

Red *12V

power

Black *Gound

*wired to the 115Vac/12DC transformer supplied with the 43502, or
separate 12Vdc supply

'CR3000 Series Datalogger

'Declare Variables and Units
Public RTD1_Res, RTD1_Cal_C
Public RTD2_Res, RTD2_Cal_C
Public 43502_Tach
Public 43502_Tach_1
Units 43502_Tach = Hz
Units 43502_Tach_1 = Hz

'Define Data Tables
DataTable (PRT_Data,1,1000)
DataInterval

(0,15,Min,1)

Average(1,RTD1_Cal_C,IEEE4,False)
Average(1,RTD2_Cal_C,IEEE4,False)
Sample

(1,43502_Tach,FP2)

Sample

(1,43502_Tach_1,FP2)

EndTable

'Main Program
BeginProg
Scan

(1,Sec,0,0)

D-3

Advertising