2 cr1000 example program 2 – Campbell Scientific AM25T 25-Channel Solid State Multiplexer User Manual
Page 16

AM25T Solid State Multiplexer
‘Main Program
BeginProg
Scan
(1,Sec,0,0)
AM25T
(TC(),25,mV2_5C,1,1,TypeT,Tref,5,4,Vx1,True,0,250,1.0,0)
CallTable
Dat15sec
CallTable
Dat5min
NextScan
EndProg
6.1.1.2 CR1000 Example Program 2
In this example, one AM25T instruction is used to measure the on-board PRT,
and another AM25T instruction is used to measure ten type T thermocouples.
Table 6-2 provides the wiring for the example.
TABLE 6-2. Wiring for CR1000 Example 2
Function AM25T
CR1000
+12 Power
12 V
12 V
Power and Shield Ground
Gnd
Gnd
Clock CLK
2
Reset RES
1
RTD Excitation
EX
EX1
Common High
HI
1H
Common Low
LO
1L
Example 2. CR1000 Program that Uses Separate AM25T Instructions to Measure the PRT
and Thermocouples
'Declare Variables and Units
Public Batt_Volt
Public RTemp_C
Public Temp_C(10)
Units Batt_Volt=Volts
Units RTemp_C=Deg C
Units Temp_C=Deg C
'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,60,Min,0)
Sample(10,Temp_C(),FP2)
EndTable
'Main Program
BeginProg
Scan(30,Sec,1,0)
'Default Datalogger Battery Voltage measurement Batt_Volt:
Battery(Batt_Volt)