B.2 edlog programming for long lead lengths – Campbell Scientific BlackGlobe Temperature Sensor for Heat Stress User Manual
Page 29

Appendix B. Edlog Programming Examples
TABLE B-2. Actual Temperature, Sensor
Resistance, and Computed Temperature
Temperature
°C
Resistance
OHMS
Output
°C
74.00 13405
73.98
76.00 12479
75.99
78.00 11625
78.01
80.00 10837
80.02
82.00 10110
82.03
84.00 9438.1
84.04
86.00 8816.9
86.03
88.00 8241.9
88.00
90.00 7709.7
89.96
92.00 7216.3
91.89
94.00 6758.9
93.80
96.00 6334.5
95.67
98.00 5940.5
97.51
100.00 5574.3
99.31
B.2 Edlog Programming for Long Lead Lengths
The following is a portion of an example CR10X program that uses the Excite-
Delay (SE) (P4) instead of the AC Half Bridge (P5).
01: Excite, Delay,Volt(SE) (P4)
1: 1
Rep
2: 3
±25 mV slow range
;On the 21X and CR7 use the 50 mV input range.
3: 1
IN Chan
;Entry depends on the datalogger SE channel used
4: 1
Excite all reps w/EXchan 3 ;Entry depends on the excitation channel used
5: 2
Delay
(units
.01sec)
6: 1000
mV Excitatio
;On the 21X and CR7 use the 2000 mV excitation
7 1
Loc [BGTemp_C ]
8: .2
Mult
;Use a multiplier of 0.1 with a 21X or CR7
9: 0
Offset
B-9