6. Specifications

Features:

• Integral switching circuit limits power consumption to measurement

cycle

• Compatible with all Campbell Scientific dataloggers (including the

CR200(X) series)

• Calibration NIST traceable

• Meets CE conformance standards

Compatibility

Dataloggers: CR200(X)

series

CR800

series

CR1000

CR3000

CR5000

CR9000(X)

CR7X

CR510

CR10(X)

CR23X

21X

6.1 Performance

6.1.1 Performance for “Standard” Range Option

Measurement Range:

600 mb to 1100 mb (hPa)

Operating Temperature Range: -40° to +60°C (-40° to +140°F)
Storage Temperature Range: -60° to +120°C (-76° to +248°F)
Proof Pressure: 1500

mb

Burst Pressure: 2000

mb

Humidity Range:

non-condensing (up to 95% RH)

Media Compatibility:

non-corrosive, non-condensing air or gas

Resolution: 0.01

mb

Total Accuracy

1

:

±0.5 mb @ 20°C

±1.0 mb @ 0° to +40°C

±1.5 mb @ -20° to +50°C

±2.0 mb @ -40° to +60°C

Linearity: ±0.4

mb

Hysteresis: ±0.05

mb

Repeatability: ±0.03

mb

Long-term Stability:

±0.1 mb per year

1

The root sum squared (RSS) of end point non-linearity, hysteresis, non-

repeatability and calibration uncertainty.

6