3 period averaging measurement example for cr23x – Campbell Scientific CS115 Barometric Pressure Sensor User Manual

Page 13

Advertising
background image

CS115 Barometric Pressure Sensor

9

;Turn off CS115
;
4: Do (P86)

1:

58*

Set Port 8* Low

5: If time is (P92)

1:

0

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

10

Set Output Flag High (Flag 0)

6: Real Time (P77)

1:

0110

Day,Hour/Minute (midnight = 0000)

7: Sample (P70)

1:

1

Reps

2:

1*

Loc [ P_mb ]

*Table 2 Program

02: 0.0000

Execution Interval (seconds)

*Table 3 Subroutines

End Program

-Input Locations-
1 P_mb

* Proper entries will vary with program and datalogger channel, and input location assignments.

5.1.3 Period Averaging Measurement Example for CR23X

EXAMPLE 3. Sample Program for CR23X

;{CR23X}
;
*Table 1 Program

01: 1

Execution Interval (seconds)

;Turn on CS115 one minute before the hour
;
1: If time is (P92)

1:

59

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

41*

Set Port 1* High

;Measure CS115 at the top of the hour
;
2: If time is (P92)

1:

0

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

30

Then Do