Appendix D. SDI-12 Sensor Support

TABLE D-2. Send Data Return Values

SDI-12

Command

Number of

Values

Returned

Values Returned

D!

7

Level [m], Level [cm], Level [ft],

Pressure [mbar], Pressure [psi],

Temperature [°C], Status

D0!

1

Level [m]; resolution: 0.001 m

D1!

1

Level [cm]; resolution: 1 cm

D2!

1

Level [ft]; resolution: 0.01 ft

D3!

1

Pressure [mbar]; resolution: 0.01 mbar

D4!

1

Pressure [psi]; resolution: 0.001 psi

D5!

1

Temperature [°C]; resolution: 0.1 °C

D6!

1

Status

D.2 SDI-12 Transparent Mode

System operators can manually interrogate and enter settings in probes using

transparent mode. Transparent mode is useful in troubleshooting SDI-12

systems because it allows direct communication with probes. Datalogger

security may need to be unlocked before transparent mode can be activated.

Transparent mode is entered while the PC is in telecommunications with the

datalogger through a terminal emulator program. It is easily accessed through

Campbell Scientific datalogger support software, but is also accessible with

terminal emulator programs such as Windows HyperTerminal. Datalogger

keyboards and displays cannot be used.

The terminal emulator is accessed by navigating to the Datalogger menu in

PC200W, the Tools menu in PC400, or the Datalogger menu in the Connect

screen of LoggerNet.

The following examples show how to use LoggerNet software to enter

transparent mode and change the SDI-12 address of a CS470/CS471 sensor.

The same steps are used to enter transparent mode with PC200W and PC400

software after accessing the terminal emulator as previously described.

D.2.1 CR200(X) Series Datalogger Example

1. Connect a single CS470/CS471 to the CR200(X) (see TABLE 7-2)

2. In the LoggerNet Connect screen, navigate to the

Datalogger menu and

select

Terminal Emulator. The terminal emulator window will open. In

the

Select Device menu, located in the lower left-hand side of the window,

select the CR200Series station.

3. Click on the

Open Terminal button.