Appendix b. sensorex’s model do6200/t, B.1 do6200/t specifications, B.1 do6200/t – Campbell Scientific CS511 Dissolved Oxygen Probe User Manual
Page 31: Appendix b, Appendix b)

Appendix B. Sensorex’s Model
DO6200/T
Prior to June 2008, Campbell Scientific’s CS511 was Sensorex’s model
DO6200/T (see FIGURE B-1) instead of Sensorex’s model DO6400/T.
Programming, wiring, and some specifications are the same for these two
sensors. However, they look different and use different accessories.
FIGURE B-1. Sensorex’s model DO6200/T
B.1 DO6200/T Specifications
Principle of Measurement:
Membrane covered, galvanic oxygen probe
Output Signal:
1.65 mV
± 0.45 mV per mg/l
Accuracy:
Better than
± 2 % of reading ± 1 digit when
calibration temperature equals measuring
temperature
± 5°C
Output Impedance:
10 k
Ω nominal
Response Time:
After equilibration, 2 min. for 90% of final
value
Materials of Construction:
Probe
body:
O-rings:
Membrane:
Delrin
Membrane O-ring = Buna N
Body Seal O-ring = Viton
0.05 mm Teflon
B-1