NR01 Four-Component Net Radiation Sensor

3.3 IR01 Pyrgeometer Specifications

TABLE 3.3-1. Specifications of IR01

IR01 Specifications
Overall classification
according to ISO / WMO

Not applicable

Response time for 95 %
response

18 s

Window heating offset
(response to 1000 W/m

2

net

thermal radiation)

< 15 W/m

2

Zero offset b (response to 5
k/h change in ambient
temperature)

<4 W/m

2

Non-Stability

< 1% change per year

Non-Linearity

< +/- 2.5%

Field of view

150 degrees

Spectral selectivity

Not specified

Temperature response
(within an interval of 50
degrees C)

Within 6% (-10 to +40 degrees C)

Tilt response

Within +/- 2%

IR01 Additional Measurement Specifications
Sensitivity

5 – 15 μV/Wm

-2

Expected voltage output

Meteorological application: -5 to + 5 mV

Sensor resistance

Between 100 and 400 Ohms

Power required

Zero (passive sensor)

Range

To 1000 Wm

-2

Spectral range

4500 to 50000 nm (50% transmission points)

Required programming

Φ = U / E (in case of net radiation only) Φ = (U /
E) + 5.67.10

-8

T

4

(absolute radiation), with T

from Pt100 measurement

Expected accuracy for daily
sums

+/- 10%

Calibration

Calibration traceability

International temperature standard

13