Supported commands, Sdi-12 commands supported – Campbell Scientific Sentek EnviroSCAN EasyAG SDI-12 Manual v 3.4 User Manual

Page 20

Advertising
background image

EnviroSMART & EasyAG SDI-12 Probe Manual

Copyright © 1991 – 2003 Sentek Pty Ltd All rights reserved

Page 15

Supported Commands

SDI-12 Commands Supported

The following commands are supported by the EnviroSMART and EasyAG SDI-12 probe interfaces:

Command

Name

Response

a!

Acknowledge
Active

a<CR><LF>
The probe interface address

aI!

Send
Identification

allccccccccmmmmmmvvvxxxxxxxxxxxx<CR><LF>
Identification information

aAb!

Change Address

a<CR><LF>
The probe interface address

aM!

Start
Measurement

atttn<CR><LF>
Delay (ttt) in seconds and number of
values (n) up to 9

aMn!

Start
Measurement
n in range “1”
- “3”

atttn<CR><LF>
Delay (ttt) in seconds and number of
values (n) up to 9
Note: A response of a0000<CR><LF> is
provided indicating all values are
served by previous Start Measurement
commands (aMn!)

aC!

Concurrent
Measurement

atttnn<CR><LF>
Delay (ttt) and number of values (nn) up
to 16

aCn!

Concurrent
Measurement
n in range “1”
- “9”

atttnn<CR><LF>
Delay (ttt) and number of values (nn) up
to 16
Note: A response of a0000<CR><LF> is
provided indicating all values are
served by previous Start Concurrent
Measurement (aCn!) commands

aDn!

Send Data
n in range “0”
- “9”

Refer to “Data Reading” section

aV!

Start
verification

a0000<CR><LF>

aRn!

Continuous
Measurement

a<CR><LF>
Not supported
Probe interface will respond with its
address

Advertising