Wiring, Datalogger programming – Campbell Scientific SP-LITE Solar Radiation Sensor User Manual

Page 11

Advertising
background image

SP-LITE Silicon Pyranometer

7

4. Wiring

The SP-LITE can be measured with a differential or single-ended channel on
the datalogger; a differential channel is recommended.

FIGURE 8. SP-LITE to Datalogger Connections

TABLE 1. Datalogger Connections for Differential Measurement

Function

Color

CR10X,CR510

CR23X

21X/CR7

CR1000

Radiation Signal

White

Differential H

Differential H

Differential H

Differential H

Signal Reference

Green

Differential L

Differential L

Differential L

Differential L

Jumper to

AG

Shield

Clear

G

TABLE 2. Datalogger Connections for Single-Ended Measurement

Function

Color

CR10X,CR510

CR23X

21X/CR7

CR1000

Radiation Signal

White

S.E. Channel

S.E. Channel

S.E. Channel

S.E. Channel

Signal Reference

Green

AG

Shield

Clear

G

5. Datalogger Programming

The SP-LITE outputs a low-level voltage ranging from 0 to a maximum of
about 15mV. A differential voltage measurement (Instruction 2, VoltDiff) is
recommended because it has better noise rejection than a single-ended
measurement. If a differential channel is not available, a single-ended
measurement (Instruction 1, VoltSE) can be used. The acceptability of a single-
ended measurement can be determined by simply comparing the results of
single-ended and differential measurements made under the same conditions.

Radiation Signal

White

Signal Reference

Green

Shield

Clear

(User supplied jumper wire for differential measurement only)

Advertising