SS100 Sensor Simulator

1. General

The SS100 is a device used when the Campbell Scientific CR1000/LoggerNet
Training Course is taught out-of-house. The SS100 is powered by a nominal
12VDC power source, typically the datalogger itself, and provides several
outputs typical of those from sensors commonly measured by Campbell
Scientific dataloggers.

The Training Course instructors may use the SS100 in a variety of simulations
of their own choosing or use the analogies in this manual.

2. Specifications

Supply Voltage
12VDC

Analog Outputs
0-2.2

VDC


Pulse Outputs

High Frequency Pulse: 0-5 VDC, 3-140 Hz, non-linear

Low Level AC: 1 VDC, 3-140 Hz, non-linear


Bridge Measurement

Full bridge circuit

Fixed resistors 5 Kohm

Variable resistor 10 Kohm


Switch Closure

Momentary push button switch, normally open

Approximate duration: 135 ms

3. Wind Speed and Direction Simulation

Sensor: Wind Speed and Direction

Wind Direction (Use SS100 Analog Output)
Range:

0-360°

Accuracy:

+/-

3%

Resolution:

Wind Direction Voltage: 0-2200 mVDC


Wiring

SS100

CR1000

+ SE1
G


Multiplier and offset calculation:

Multiplier = 360 degrees / 2200 mV

Offset

=

0

1