Performing a Quality Control Check

1. Open test strip vial and remove test strip.

Replace vial cap.

2. Within 3 minutes, insert

test strip (yellow window
facing up) into test strip
slot. The meter turns on
automatically.

Check that the code
number on the meter
matches the code on the
vial of test strips you are
using. The test strip symbol
stops flashing and a blood
drop symbol appears on the display.

3. Press and release the right arrow button once to

select a L1 (Level 1) control or twice to select a
L2 (Level 2) control.

Note: Results marked as L1 or L2 do not appear
in memory or averages.

4. Hold the control bottle

horizontally with the tip
pointed directly at the right
edge of the strip.