Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual
Page 25
 
Sampling Problems 
Airflow is restricted during the 
sensor regeneration cycle, causing 
possible permanent damage. 
Kinks and crimps in the
Tygon tubing.
Periodically check the Tygon tubing
inside the instrument. Refer to page 
20. 
High erratic results.
Internal hydrogen sulfide
contamination.
1. Install zero air filter in intake
and tighten intake nut. Press 
SAMPLE button. After three 
samples, if readings are over 
0.003 ppm, replace fritware 
filter, intake filter disk, and 
Tygon tubing. Refer to page 19.
2. Perform sensor regeneration
with the zero air filter in intake. 
Refer to page 12. Retest if 
necessary. Replace scrubber 
filters and Tygon™ tubing. 
Refer to page 20. 
High/erratic results
Intake and internal filters
may become clogged and 
need replacement when 
sampling in a dusty or humid 
area. 
1. Open instrument and check for
pinched, crimped or 
disconnected internal tubing. 
2. In extreme conditions, an
additional particle filter may be 
installed on the intake. 
High/erratic results 
Readings vary more than 0.003 ppm 
when in survey mode.
Loose connections to gold
film sensor.
Place a zero air filter into the intake.
Place the instrument in survey 
mode. Move the unit as samples 
are being taken. Call AZI 
Customer Service at 800-528-7411 
or 602-470-1414 for assistance. 
Low response or erratic readings
after a long period of non-use.
May need a second
regeneration cycle.
1. Wait 30 minutes and perform
another sensor regeneration.
2. Test with FTM. Refer to page
3. If still unresponsive,
False readings, may go to .8.8.8 or
.L.L.L.
Extremely cold or extremely
warm air sampled into unit.
If sampling under these conditions,
install zero air filter in intake. 
Sample until display reads 0.003 
ppm or less. This equilibrates 
sensor temperature with the 
temperature of the sample air 
stream. Remove filter and take 
samples. 
AZI Customer Service 800-528-7411 or 602-470-1414
Page 25 of 50