GENERAL
INTENDED USE
Designed to meet the demanding requirements for scientific and laboratory research.
Advanced engineered design incorporates the latest in cabinet, refrigeration, temperature
control and monitoring features. Provides energy efficient, convenient, safe and reliable
performance for optimal temperature environments necessary for a wide range of plant
science, life science, pharmacy, biological, medical, clinical and industrial applications.
TECHNICAL SPECIFICATION
Voltage supply:
115V 60HZ 1 phase
Maximum fuse size: 20A
Total amp draw:
13.6A
Max set point :
44 C degrees
INSPECTION
When the equipment is received, all items should be carefully checked against the bill of
lading to insure all crates and cartons have been received. All units should be inspected for
concealed damage by uncrating the units immediately. If any damage is found, it should be
reported to the carrier at once, and a claim should be filed with the carrier. This equipment
has been inspected and tested in the manufacturing facility and has been crated in
accordance with transportation rules and guidelines. Manufacturer is not responsible for
freight loss or damage.
LOCATION
The cabinet should be located within reach of an outlet that has 115V 60HZ 20A with a
protective earth ground. The outlet should be easily accessible when installation is
complete. The refrigeration system located at the top of the cabinet requires free air
access for proper operation. Allow a minimum four-inch clearance on the top, rear, and
sides of the cabinet. The cabinet should also be leveled when it is placed in its permanent
location. Do not stack items on top of the unit. Vibration during shipping and handling may
loosen mechanical connections. Check all connections during installation. Check all
wiring, piping and fasteners. This unit has been designed for 75 deg with 50%RH at up to
2000 Meter elevation. If equipment is not used for its designed manner specified by the
manufacturer, the protection provided by the equipment may be impaired.
WARNINGS AND CAUTIONS
• Do not modify cabinet construction or associated equipment assemblies.
• Do not remove labeling or information supplied with the unit.
Warning: Electric Shock Hazard
.
Do Not Remove top electrical cover.
Contact a qualified service representative.