ELECTRONIC CALIBRATION (CIRCUIT ONLY) -

See Section V.B.1.

a. Set Point Control Calibration

1. Ensure power is OFF.

2. Using a standard slot screwdriver remove the four (4)

screws on the clear cover and remove cover.

3. The front panel is held in place with hook & loop strips

on each side. By carefully lifting at the slot on the top,

the front panel will pull toward you. Do not pull more than

about 8 inches/20CM.

DO NOT pull on control knob(s).

4. Rotate the front panel down so that the back side is

facing you. The Circuit Board is mounted to the back of

the front panel.

5. Turn power ON.

Caution: When handling CB, DO NOT touch component side.

Electrical shock may occur.

6. Press and hold the EC switch.

7. Rotate Set Point control from “0” toward EC. At EC the

relay should click, and the LED should change from RED

to GREEN. If not, set Set Point control to EC and adjust

the CALibration control.

Models with 3 1/2 digit LCD.

The display should indicate 2.50.

If not, adjust the DISplay calibration control.

8. Turn power OFF.

9. Re-install front panel and cover as described in

“REASSEMBLY”.

10. To operate, turn power ON.

b. 0-10VDC Recorder Output Calibration

See section V.B.1.

USING STANDARD SOLUTIONS - Section V.B.2.

The BEST method of verifying and recalibrating your CTCII

Chemical Treatment Controller is with NIST traceable Standard

Solution (available from the Myron L Company). Because it

includes the sensor, the entire system is calibrated.

The following procedure describes the easiest method for

standard solution calibration. We recommended using Myron L

442-3000 Standard Solution or equivalent. This standard solution

will calibrate either Conductivity µS (at 3900 µS) or TDS ppm (at

3000 ppm).

1. Rinse thoroughly, and fill a clean glass beaker with

standard solution.

2. Place sensor in beaker of standard solution. The level of

standard solution must be high enough to cover at least

1/2” above the electrodes.

3. Slowly shake the sensor to remove air bubbles from the

electrodes.

4. Allow 5-10 minutes for temperature to equilibrate. For

the quickest and the best results, both the sensor and

solution should be at the same temperature.

5. Turn power ON.

6. Rotated the Set Point control to 3900 for Conductivity

models or 3000 for TDS models. The relay should click

and the LED will change from GREEN to RED.

If the set point is different, adjust CALibration control

on the main circuit board until the reading matches the

solution.

Models with display (standard 3 1/2 digit).

The display should match the value of the standard

solution, 3900 µS (displayed as 3.90) or 3000 ppm

(displayed as 3.00).

If the reading is different, adjust DISplay calibration

control on the main circuit board until the reading

matches the solution.

7. After adjustment, turn power OFF.

8. Re-install front panel and cover as described below in

“REASSEMBLY”.

9. To operate, turn power ON.

TRANSFER STANDARD METHOD - See Section V.B.3.

CTCII MAIN CIRCUIT BOARD

Figure V.A.1

115/230 VAC

SELECT SWITCH

100mA Slow Blow (T.10A)

115 & 230 VAC (V~)

Wickman Fuse 5 Amp

(#3741500041 or equiv)

115 VAC(V~)

SCALE SELECT

ELECTRONIC

CALIBRATION

MAIN

CALIBRATION

CONTROL

DISPLAY

CALIBRATION

CONTROL

(FACTORY SET)

OPTIONAL

LIMIT

TIMER

CONTROL

MODULE

OPTIONAL

4-20 mA

OUTPUT

MODULE

DIS

CAL

TDS

COND

EC

CTCII

DISPLAY

TRANSFORMER

Yel

CAP

CAP

C4AM

LTM

HYST

Flow Switch

Jumper -

Remove if

flow switch

installed.

SWITCH

FUSE

RELAY

RELAY

BLEED FEED

FEED

PUMP

BLEED

VALVE

POWER

FLOW

SWITCH

SENSOR

0-10VDC

FS

BK WTRD GN NU R- R+

PWR

C GD

PWR

C GD

PWR

C GD

SET POINT

HYSTERESIS

RIGHT INC

LEFT DEC