FEaTuRES and SPEciFicaTiOnS

A. Features

ORP-based FC

E

Free Chlorine measurement; displays as ppm

concentration

Alkalinity and Hardness Conductometric Titrations

Langelier Saturation Index (LSI) Titrations and Calculator

Superior resolution 4 digit LCD displays full 9999 µS/ppm

Cond/MIN/SALT/TDS Accuracy of

±1% of READINg / ±0.1% at

calibration point

All electrodes are internal for maximum protection

Improved 4-electrode sensor technology

Waterproof to 1 meter/3 feet

Autoranging conductivity/TDS

Prompts for easy pH calibration

Factory calibrations stored in microprocessor

3 conductivity/TDS solution conversions preprogrammed into

microprocessor

Real Time Clock with Time and Date

Data Logging with TIME and DATE in memory

Memory stores 100 readings

User adjustable timeout “

auto OFF

Bluetooth® wireless download capability with optional bluDock™

accessory package

B. general Specifications

Display

4 Digit LCD

Dimensions (LxWxH)

196 x 68 x 64 mm/

7.7 x 2.7 x 2.5 in.

Weight

352 g/12.4 oz.

Case Material

VALOX*

Cond/Res/TDS Cell Material

VALOX*

Cond/TDS Electrodes (4)

316 Stainless Steel

Cond/Res/TDS Cell Capacity

5 ml/0.17 oz.

pH/ORP Sensor Well Capacity

1,2 ml/0.04 oz.

Power

9V Alkaline Battery

Battery Life

>100 Hours/5000 Readings

Operating/Storage Temperature

0-55°C/32-131°F

Protection Ratings

IP67/NEMA 6 (waterproof to

1 meter/3 feet)

EMI/EMC Ratings

EN61326-1: 2006 + Annex A: 2008

(hand-held devices)

(Conformité Européenne)

CISPR 11: 2003

IEC 61000-4-2: 2001 and,

IEC 61000-4-3: 2002

* ™ SABIC Innovative Plastics IP BV