Myron L 6P (includes Addendum 10-01) User Manual

Page 12

Advertising
background image

for ORP are highly reactive and potentially hazardous, your Ultrameter
has an electronic ORP calibration. This causes the zero point on the
reference electrode to be set whenever pH 7 calibration is done.

7. Temperature Calibration

Temperature calibration is not necessary in the Ultrameter.

VII.

MEMORY

This feature allows up to 20 readings with their temperatures to be stored
simultaneously for later recall.

A. Memory Storage

1.

While displaying a measurement, press to record the
displayed value.

2.

MEMORY

” will appear and

the temperature display will be
momentarily replaced by a
number (1-20) showing the
position of the record. Figure
9 shows a reading of 1806 µS
stored in memory record #4.

B. Memory Recall

1.

Press one of the five measurement keys.

2.

Press , “

MEMORY

” will appear, and the display will show

the last record stored.

3.

Press the or

to scroll to the record location desired

(the temperature display alternates between temperature
recorded and location number).

4.

Press any measurement key to leave memory recall or allow to
automatically turn off.

C. Clearing a Record/Memory Clear
After recalling a certain record location, press

to clear that

memory. This space will be the place for the next memory record,
unless you scroll to another position before ending the recall
sequence. The next memory stored will go into the next highest

available memory location.

Example:

You have locations 1-7 filled. You want to clear the conductivity

reading stored in record location #

3

and replace it with a pH reading.

1.

Press and scroll to location #

3

.

2.

Press to clear old record #

3

.

3.

Fill pH/ORP sensor well with sample.

4.

Press to measure sample and press

to store

reading in location #

3

.

5.

The next memory stored will go into location #

8

.

6.

To clear

all

records: After

pressing , scroll down

to “

CL

r

ALL

” in measurement

and temperature area
(see Figure 10).

7.

Press . All records will be cleared.

VIII.

CHANGING from CENTIGRADE to FAHRENHEIT

1.

Press .

2.

Press to display the stored memory records.

3.

Press repeatedly until you pass the memory “

CL

r

ALL

location. The display will show a “

C

” or “

F

(see Figures 11 & 12).

20

21

MR

COND

MR

CAL

MCLR

pH

Figure 10

MR

M S

MR

CAL

MCLR

CAL

MCLR

MR

M S

MR

M S

Figure 9

COND

µ S

°C

MEMORY

4 4 2

MEMORY