Myron L 6PFCE and 4P User Manual

Page 14

Advertising
background image

3.

Press

or

.

4.

Take reading. A display of [- - - -] indicates an overrange condition.

B. Measuring Resistivity

Resistivity is for low conductivity solutions. In a cell cup the value may

drift from trace contaminants or absorption from atmospheric gasses, so

measuring a flowing sample is recommended.

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Hold instrument at 30° angle (cup sloping downward).

3.

Let sample flow continuously into conductivity cell with no aeration.

4.

Press

key; use best reading.

NOTE: If reading is lower than 10 kilohms display will be dashes:

[ - - - - ]. Use Conductivity.

C. Measuring pH (6PFC

E

)

1.

Remove protective cap by rotating while grasping and pulling up.

2.

Rinse pH/ORP sensor well and conductivity cell cup 3 times

with sample to be measured. Shake out each sample to

remove any residual liquid.

3.

Refill both sensor well and cell cup with sample.

4.

Press

.

5.

Note value displayed.

6.

IMPORTANT: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap.

If Myron L pH Sensor Storage Solution is unavailable, use a

strong KCl solution, a pH 4 buffer, or a saturated solution of

table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP, pg.

34).

Do not allow pH/ORP sensor to dry out.

D. Measuring ORP

The Ultrameter II features the ability to measure the activity of oxidizing or

reducing chemicals in solution as ORP mV. The instrument also includes

an innovative Free Chlorine Equivalent (FC

E

) feature (Measuring Free

Chlorine Using FC

E

, pg. 12) that uses ORP and pH to measure free

available chlorine (FAC) concentration in ppm. ORP mV and ppm of

free available chlorine (FAC) are the two most commonly used sanitizer

units of measure in water quality management.