First steps – BrandTech Scientific seripettor User Manual

Page 35

Advertising
background image

b

a

!

30 mm

English

35

Before using the instrument for the first time, ensure it is rinsed
carefully and discard the first few samples dispensed. Avoid splashes.

Note:

Never press down the piston when the discharge tube is closed with
the stopper/screw cap. Avoid splashing the reagent! The reagent can
drip out from the discharge tube and stopper cap.

Warning:

Assembly (cont.)

6.

Priming the

instrument

a) Hold the discharge tube and loosen

the stopper cap/screw cap (fig. a).

b) To avoid splashes, hold the discharge

tube orifice on the inner wall of a

suitable receiving vessel.
Allow the piston to rise up approx.

30 mm and push it down rapidly until
the lower stop (fig. b).
Repeat this procedure approximately
5 times until the discharge tube is
bubble-free. A few bubbles up to
1 mm in size are permissible.

First Steps